분광측색계 CM-2600d/2500d
Features
-
Numerical Gloss Control, 상대 Gloss 값 표시
-
휴대가 용이한 Body와 밝은 finder를 이용하여 정확한 측정 부위 설정 가능
-
전지를 이용한 구동 방식, 경량, 조그 방식의 다이얼을 이용한 조건에 구애 받지 않는 측정 가능
-
Numerical 기술로 기계적 구동을 최소화 하여 고신뢰성 확보 및 유지보수의 최소화 가능
-
Numerical UV Control(Only CM-2600d)
-
ISO, JIS, DIN, CIE 등 국제산업규격 준수 및 L*a*b*, Yxy, Munsell 및 CMC 등 색채계 표시
-
3mm, 8mm 측정경 교환 가능(CM-2600d의 경우)
Specifications
Model
Measuring geometry
Sphere size
Detector
Wavelength range
Wavelength pitch
Half bandwidth
Photometric range
Light source
Measurement time
Minimum measurement interval
Battery performance
Illumination/
measurement areas
Repeatability
Inter-instrument agreement
UV admustment
Measurement mode
Interface
Observer
Illuminant
Displayed data
Color space
Color space/colorimetric data
Storable data
Tolerance judgement
Power
Operation temperature/humidity range
Storage temperature/humidity range
Dimensions
Weight
CM-2600d/2500d
d/8(diffused illumination, 8-degree viewing angle), equipped with simultaneous measurement of SCI(specular component included)/SCE(specular component excluded). (Conforms to DIN 5033 JIS Z 8722 Condition C, ISO 7724/1, CIE No.15, ASTM E1164)
Φ52mm
Silicon photodiode array (dual 40 elements)
360 to 740 nm
10nm
Approx. 10nm
0 to 175%(Reflectance or transmittance), Output/display resolution: 0.01%
3 pulsed xenon lamps(2 xenon lamps for CM-2500d)
Approx. 1.5 seconds (approx. 2 seconds for fluorescent measurement)
3 seconds for SCI/SCE(4 seconds for fluorescent measurement)
(Simultaneous evaluation of SCI/SCE is possible by a single measurement)
Approx. 1000 times at 10-second intervals (when alkaline batteries used
MAV: Ø8mm/Ø11mm
SAV: Ø3mm/Ø6mm(Selectable between MAV and SAV, CM-2600d only)
Spectral Reflectance:Standard deviation within 0.1%(360 to 380nm within 0.2%)
Colorimetric Value:Standard deviation within ΔE*ab 0.04(When white calibration plate is measured 30 times at 10-second intervals after white calibration was performed)
Delta E*ab within 0.2(MAV/SCI)
(Based on 12 BCRA Series II color tiles compared to values measured with master body)
Instantaneous adjustment(without mechanical moving parts) with UV400nm cut filter(CM-2600d only)
Single/averaging(auto mode: 3, 5, 8 flashes/manual mode)
RS-232C
2° Standard Observer or 10° Standard Observer
A, C, D50, D65, F2, F6, F7, F8, F10, F11, F12
(simultaneous evaluation is possible using two light sources)
Spectral value/graph, colorimetric value, color difference value/graph, PASS / FAIL result
L*a*b*, L*C*h, Hunter Lab, Yxy, XYZ, Munsell, and color difference in these spaces (except for Munsell)
L*a*b, L*C*h, CMC (1:1), CMC (2:1), CIE94, Hunter Lab, DIN99, XYZ, Yxy, Munsell, MI, WI (ASTM E313), WI / Tint (CIE, Ganz/Griesser)
YI (ASTM E313/ASTM D1925), ISO Brightness (ISO 2470),Density status A/T
1700 (SCI/SCE as a set)
Tolerance for color difference(both box and elliptical tolerances can be set)
AA-size battery (x4), Rechargeable batteries, AC adapter
5 to 40°C; less than 80% RH(no condensation); Installation category: 2; Pollution degree: 2
0 to 45°C; less than 80% RH(no condensation)
69 x 96 x 193mm
Approx. 670g(without batteries